Next: 4 Putting the GC/MS
Up: 3 Warm Start: Starting
Previous: 3.1 GC
Contents
- Check oil level in Roughing Pump. If the oil is low or badly discolored,
add or change the oil.
- Be sure Turbo Pump is spinning full speed as indicated by the meter
on the front of the MS. Do NOT proceed if the Turbo Pump is not at
full speed.
- With the Turbo Pump at full speed, check the high vacuum with the
Ion Gauge.
- Turn Power to the Gauge Controller ON.
- Set the Multiplier to the lowest setting, such 3 (10
torr) on some units or 4 (10
) on others.
- Turn the Filament on the Gauge Controller ON. You should be able to
see the filament in the Gauge Tube glowing.
- If the meter needle goes off scale and/or the filament turns off,
there is an air leak. This leak should be repaired before continuing.
See Section Vacuum Leak Detection and Repair/High Vacuum Leaks.
- Insure that the high vacuum pressure is 1.0*10
torr or lower. The lower the better. If the instrument is being Cold
Started, this may take considerable time; HP recommends a minimum
of 3 hours equilibration time before actually using the instrument
after a Cold Start.
Note: It is a good idea to record in your GC/MS Instrument logbook
what this daily pressure reading is, so that you have a record of
'normal' operating pressure for your instrument. This can greatly
simplify troubleshooting vacuum problems.
- Reset the Gauge Controller Multiplier to its lowest setting.
- Very slowly, and while watching the Gauge Meter, open the calibration
gas valve. You may see the needle rapidly increase then decrease.
After this jump, it is okay to open the valve the rest of the way.
Make sure the pressure goes back below 1*10
torr
before proceeding.
- Turn OFF the Ion Gauge Filament and Power.
- Launch the ChemStation Software (if not already running).
- If doing a Cold Start or if you suspect a small leak (the Step 4 reading
higher than normal but still in the 10
torr range
or lower), run the Air/Water Check utility. The m/z 28 and 32 signals
should be smaller than m/z 18. The m/z 18 should be smaller than the
PFTBA m/z 69.
If the Air/Water Check does not pass, it could be that the system
simply has not pumped long enough after a Cold Start, or there may
be a small air leak.
- Perform an MS Autotune if needed. HP does not recommend autotuning
their instruments daily, unless there is some compelling reason to
do so for instruments used continuously or in Stand-By mode when not
in use. Autotuning every other day or once per week is generally adequate.
An Autotune should be always be done after a Cold Start.
The Autotune printout should be saved in an MS Autotune Logbook for
your instrument. The operator should examine the Autotune printout
and compare the new Autotune parameters to the last Autotune parameters:
- Dramatic changes of any parameter are often an indication of an instrument
needing maintenance.
- The EM Voltage (usually somewhere around 2000 Volts) will gradually
increase as the EM ages.
- When the EM voltage gets to around 2500 Volts, it is a good idea to
order an EM.
- Mass axis calibration should be within about +/- 0.2 amu for the three
PFTBA peaks m/z 69, 219 and 502. If the masses of these three peaks
deviate by more than 0.2 amu, rerun the Autotune to see if the mass
calibration improves.
- Signal widths at half height for each of the three tune peaks should
be about 0.55 - 0.60 for the default ChemStation Autotune routine.
If any of the peak half widths are less than 0.55 or greater than
0.60, rerun the Autotune to see if the widths improve.
- The m/z 219 should generally be about 40-60% of the m/z 69 signal.
If it is not, rerun the Autotune to see if the signal ratio improves.
- The m/z 502 should generally be about 3% (5% is okay, but smaller
than about 3% may be an indicator of an instrument that requires
maintenance) of the m/z 69 signal. If it is not, rerun the Autotune
to see if the signal ratio improves.
- If the MS has passed the Autotune, close the calibration valve.
Next: 4 Putting the GC/MS
Up: 3 Warm Start: Starting
Previous: 3.1 GC
Contents
John S. Riley, DSB Scientific Consulting