Some mass spectrometers have more than one mass analyzer, and two
is most common. If the two analyzers are of the same type, such as
two magnetic sector analyzers, it is called a tandem instrument. On
the other hand, if the analyzers are different types, the instrument
is called a tandem hybrid. Generally, an rf only quadrupole (no offset,
so no mass filtering) is placed between the analyzers and doubles
as a collision cell. The technique of using mutiple analyzers is called
MS/MS (for two analyzers), MS/MS/MS (for three), etc. Collision gas
may be added between the cells to induce ion-atom or ion-molecule
chemical decompositions.
Instrument geometries are abbreviated to communicate the analyzers used in a given experiment. If it does not matter in a particular context what type a given analyzer is, one may see MS1 (for the first analyzer), MS2 (for the second), etc. If, on the other hand, one needs to know the type of analyzer to draw the proper conclusions, the following shorthand can be used::
It is also worth noting that the quadrupole ion trap provides MS/MS/MS/MS/MS, etc capability in one small instrument with formally only one analyzer. This is accomplished by adding, removing, adding, removing (etc) the offset voltage that effectively alternates the trap between a mass analyzer and an rf only (no mass discrimination) trap. Collision gas can optionally be inserted during an rf only phase and pumped out for the next mass analysis phase to induce decomposition of ions selected in the previous mass analysis step.